Luminos dRF Max, model no. 10762471 - Product Usage: Luminos dRF Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.
Recall
- Recall Number
- Z-1337-2020
- Event Number
- 84815
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 8, 2020
- Terminated
- June 21, 2022
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Luminos dRF Max, model no. 10762471 - Product Usage: Luminos dRF Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.
A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis.
On January 8, 2020, the firm notified customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to ensure that the image displayed after acquisition is a newly acquired image and belongs to the current patient. If the error occurs and a previously acquired image is duplicated, the user should delete the image and contact Siemens Customer Service at 1-800-888-7436 to report the incident. Siemens Healthineers will correct the error with a field software update which will be available beginning January 2020. Your local Siemens Healthineers Service organization will contact you to arrange a date to perform this corrective action when the software update is available for your system. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
US Nationwide distribution.