FDA Recall Terminated

Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model SY-12120, which is included with the package.

Recall: Z-1336-2014 · Initiated March 18, 2014

Recall

Recall Number
Z-1336-2014
Event Number
67597
Firm
Energizer Personal Care
FEI Number
3007795627
Product Code
HGX
Status
Terminated
Root Cause
Component change control
Initiated
March 18, 2014
Posted
April 4, 2014
Terminated
June 10, 2016
Address
6 Research Dr, Shelton, CT, 06484-6228

Description

Playtex Nurser Deluxe Double Electric Breast Pump, Model X06578AO, powered by a 12V DC power supply, Model SY-12120, which is included with the package.

Reason

Some of the power adapters outer casings may become loose and separate, resulting in a potential for electric shock.

Action

Energizer sent and Voluntary Retail Recall Notice letter dated March 18, 2014 to retail customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to stop sale and return the affected product. The letter provides instruction to customers to return the affected product. For questions or concerns, contact Energizer Personal Care at 1-888-207-1492 or visit www.playtexproducts.com. On 9/9/2014 the firm expanded the recall to one additional production lot to include: Suspect date codes for Breast Pumps: P12315-XXXX to P13205-XXXX. Suspect adaptor date codes: 1238 to 1324 Revised Press issued 9/9/2014.

Distribution

USA Nationwide Distribution in the states of: AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, MD, MI, MN, MO, NC, NJ, NV, NY, OH, ON, OR, PA, SC, TN, TX, VA, VT, WA, WI Canada - one consignee, Energizer (recalling firm's) company store

Quantity

35,803