FDA Recall Open, Classified

MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750

Recall: Z-1335-2024 · Initiated January 30, 2024

Recall

Recall Number
Z-1335-2024
Event Number
94035
Firm
Angiodynamics, Inc.
FEI Number
1319211
Product Code
DRE
Status
Open, Classified
Root Cause
Process control
Initiated
January 30, 2024
Posted
March 21, 2024
Address
603 Queensbury Ave, Queensbury, NY, 12804-7619

Description

MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750

Reason

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Action

AngioDynamics initiated Urgent Voluntary Medical Device Recall Notification packets by Federal Express to consignees on 1/30/24. Letter states reason for recall, health risk and action to take: IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). Mark affected items with text QUARANTINE to segregate from usable stock. o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. o Notify Healthcare Providers of this possible issue that could occur during use of the Mini Stick Max device. 2. Complete and return the Reply Verification Tracking Form. " If you have any questions regarding this notice and/or if affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. " Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return), following the directions on this page and the Reply Verification Tracking Form. Email Reply Verification Tracking Form (preferred): [email protected] Fax Reply Verification Tracking Form: Attn: Mini Stick Max Recall Coordinator Fax number 1-855-273-0519 3. Package and Return the Recalled Product. " Package any product that is being returned in an appropriate shipping box. " Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. " Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: Mini Stick Max Recall Coordinator

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Quantity

4366 units