FDA Recall Terminated

Churchill Medical Systems 14"Extension Set with SmartSite Injection Item # 43000E Intended use: Intravascular Administration Set

Recall: Z-1330-2011 · Initiated November 16, 2010

Recall

Recall Number
Z-1330-2011
Event Number
57295
Firm
Churchill Medical Systems, Inc.
FEI Number
1000120555
Product Code
FPA
Status
Terminated
Root Cause
Vendor change control
Initiated
November 16, 2010
Posted
February 16, 2011
Terminated
May 1, 2012
Address
87 Venture Drive, Dover, NH, 03820-5914

Description

Churchill Medical Systems 14"Extension Set with SmartSite Injection Item # 43000E Intended use: Intravascular Administration Set

Reason

Routine testing conducted to challenge the success of sterilization was not fully completed for sterile products.

Action

The firm, Churchill Medical/Vygon, sent an "URGENT MEDICAL DEVICE RECALL" letter dated November 16, 2010, via overnight mail to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to check their stock for affected product; immediately cease use and distribution and quaratine the affected product; contact Vygon's Customer Service Department at 1-800-473-5414 for a return Merchandise Authorization number; complete and return the Recall Acknowledgement and Inventory Return Form via fax to 215-390-2019 and mail return product with form to: Vygon, 103A Park Drive, Montgomeryville, PA 18936, and if the customer is a distributor, they were to provide their customers with a copy of this notification and attached form. Note: An additional follow up communication will be provided to the customers after reconciliation of the returned inventory and all inventory will be credited or replaced. If you have any questions, please contact your local sales representative or Vygon's Customer Service Department at 1-800-473-5414, or by email at [email protected].

Distribution

Nationwide Distribution: USA including states of: CO, IA, OH, MA, MD, MO, NY, FL, SC, FL, and VA.

Quantity

19, 750 units