DigitalDiagnost C90 High Performance. radiography and fluoroscopy system

Recall: Z-1328-2022 · Initiated May 19, 2021

Recall

Recall Number
Z-1328-2022
Event Number
90490
Firm
Philips Medical Systems DMC GmbH Development And Manufacturing Ctr. Rontgenstr. 24-26 Hamburg Germany
FEI Number
3003768251
Product Code
KPR
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
May 19, 2021
Posted
July 12, 2022

Description

DigitalDiagnost C90 High Performance. radiography and fluoroscopy system

Reason

Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.

Action

Philips sent an Electronic Product Radiation Defect notification letter dated May 25, 2022 to affected customers. Philips will schedule an appointment with customers to install the software update. Philips will, without charge bring the product into compliance. Customers were instructed to complete and return the attached response form to Philips. For further information or support concerning this issue, please contact Philips Customer Care Solutions Center: 1-800-722-9377 and reference FCO71200223 for impacted DigitalDiagnost systems or FCO70600109 for impacted ProxiDiagnost systems.

Distribution

US Nationwide Distribution

Quantity

140 systems in total, (Updated 1/30/2023).