FDA Recall Open, Classified

SureSigns VS4 Government Bundle, Product Number 863286. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Recall: Z-1324-2019 · Initiated April 19, 2019

Recall

Recall Number
Z-1324-2019
Event Number
82684
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
April 19, 2019
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

SureSigns VS4 Government Bundle, Product Number 863286. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Reason

System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

Action

On April 19, 2019 Philips Healthcare issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. Actions to be taken by customer: Upon receipt of this notification, promptly perform the software upgrade to each of your SureSigns VS3/VS4 monitors, as called out in the attached Pulse Rate Software-Installation Instructions. It details how to obtain and install the system software update. It also provides instructions on how to export the file that is created during software installation and instructions where to email the file back to Philips. If you need any further information or support concerning this issue, please contact your local Philips representative or call us at 1-800-722-9377.

Distribution

Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.

Quantity

366