FDA Recall Terminated

Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number: 945-603, distributed one (1) electrode per box.

Recall: Z-1323-06 · Initiated September 19, 2005

Recall

Recall Number
Z-1323-06
Event Number
35788
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
September 19, 2005
Posted
August 3, 2006
Terminated
February 24, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number: 945-603, distributed one (1) electrode per box.

Reason

Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.

Action

Following the receipt of three (3) customer complaints, on 9/14/2005, the firm issued an E-Mail notification [Management Cover] to their Service Personnel on 9/19/2005. The Management Cover instructed the service personnel to check the electrodes located at their stock and their customer locations; put all electrodes, which cannot be inserted in their call-back stock; and provide for credit to their customers through normal channels. The Service Personnel in turn, contacted their customers via telephone on or around 9/20/2005 and asked that they check their inventories for the presence of any electrodes which would not fit in the analyzer and return any such electrodes to Radiometer America for replacement

Distribution

Nationwide and Canada.

Quantity

34