VITROS Immunodiagnostic Products TSH Reagent Pack
Recall
- Recall Number
- Z-1322-2018
- Event Number
- 79508
- Firm
- Clinical Diagnostic Systems
- FEI Number
- 3003939841
- Product Code
- JLW
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 11, 2018
- Posted
- April 10, 2018
- Terminated
- September 8, 2020
- Address
- 601 Lee Rd, Ny Rochester, NY, 14652-0001
Description
VITROS Immunodiagnostic Products TSH Reagent Pack
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
Distributors instructed to disgard remaining inventory, complete Confirmation of Receipt form, and firm promised to replace product. Distributors instructed to send letter to customers.
Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)