Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes.

Recall: Z-1322-2014 · Initiated March 28, 2013

Recall

Recall Number
Z-1322-2014
Event Number
67574
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
March 28, 2013
Posted
April 2, 2014
Terminated
July 26, 2017
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227

Description

Integrity 1.1. Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes.

Reason

The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values.

Action

An Important Field Safety Notice (IFSN) has been released and distributed to all customers. Integrity 1.1 Incorrect Static Tolerances for Linac Geometric Axes, 200 01 502 032. The notice explains the issue and that the user should apply the static Tolerance Updater tool to check tolerances and adjust as required. Important Field Safety Modification Digital Accelerator FCO Ref: 200 02 502 031 Integrity R1.1 Static Tolerance Updater Tool was released 03-2013. Elekta has released Integrity version 1.2 which contains the permanent fix for this issue. The permanent solution fixes the error that made the digital accelerator protection functions use the value loaded from the calibration database. The solution now ensures the protection functions use the values the digital accelerator calculates. Important Field Safety Notice 200 01 502 035 has been released to inform customers the solution is available. Important Field Safety Modification 200 02 502 036 will track the installation of the software by Elekta service engineers.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CT, FL, GA, KY, LA, MD, MO, NM, NY, NC, OH, OK, PA, TN, TX, and WA and the countries of Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.

Quantity

498