FDA Recall Terminated

Hematology Diluent M-53D used with BC-5390 Hematology Analyzer; Part Number: 105-007867-00 (20L container) The M-53D diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-5390 Hematology Analyzer.

Recall: Z-1321-2018 · Initiated January 24, 2018

Recall

Recall Number
Z-1321-2018
Event Number
79556
Firm
Mindray DS USA, Inc. dba Mindray North America
FEI Number
2221819
Product Code
GIF
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
January 24, 2018
Terminated
April 19, 2019
Address
800 Macarthur Blvd, Mahwah, NJ, 07430-2001

Description

Hematology Diluent M-53D used with BC-5390 Hematology Analyzer; Part Number: 105-007867-00 (20L container) The M-53D diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-5390 Hematology Analyzer.

Reason

Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzer may cause the system to produce an elevated platelet background count.

Action

Letters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.

Distribution

US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .

Quantity

496