Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L container) The M-30D Diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-3200 and BC-3600 Hematology Analyzer.
Recall
- Recall Number
- Z-1320-2018
- Event Number
- 79556
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- FEI Number
- 2221819
- Product Code
- GIF
- Status
- Terminated
- Root Cause
- Material/Component Contamination
- Initiated
- January 24, 2018
- Terminated
- April 19, 2019
- Address
- 800 Macarthur Blvd, Mahwah, NJ, 07430-2001
Description
Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L container) The M-30D Diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-3200 and BC-3600 Hematology Analyzer.
Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet background count.
Letters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.
US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .