FDA Recall Terminated

Gomco Style Circumcision Clamp Trays is an instrument used to compress the foreskin of the penis during circumcision of a male infant.

Recall: Z-1320-2016 · Initiated February 29, 2016

Recall

Recall Number
Z-1320-2016
Event Number
73497
Firm
Cardinal Health
FEI Number
3001236905
Product Code
OHG
Status
Terminated
Root Cause
Labeling Change Control
Initiated
February 29, 2016
Terminated
March 2, 2018
Address
1430 Waukegan Rd, Mc Gaw Park, IL, 60085-6726

Description

Gomco Style Circumcision Clamp Trays is an instrument used to compress the foreskin of the penis during circumcision of a male infant.

Reason

Two lots of Gomco Style Circumcision Clamp Trays, SSI-0034 and 10-1100 contained a 1.3cm circumcision clamp. They both were labelled as containing a 1.1cm clamp.

Action

Cardinal Health sent an URGENT: PRODUCT RECALL letter dated February 29, 2016, to affected consignees explaining the reason for the product recall and actions that should be taken by the end user. These included the following actions: 1. STOP using affected product. 2. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652- 9648, whether or not you have affected product, as Cardinal Health is required to confirm receipt of this notification from you, and to prevent further notices. 3. NOTIFY any customers to whom you may have distributed product affected by this recall. 4. CONTACT the appropriate Customer Service group to arrange for return of any affected product you may have: " Hospital800.964.5227 " Distributor800.635.6021 The firm asks that consignees who have question or requests of special assistance relating to the recall should contact Cardinal Health Customer Advocacy at 800.292.9332. For questions regarding this recall call 847-473-1500.

Distribution

Nationwide Distribution including AR, KS, WV, , NC, TX, AL, MS and VA.

Quantity

SSI-0034: 380 units and 10-1100: 60 units