FDA Recall Terminated

Terumo Advanced Perfusion System 1 Base, 100/120V, Catalog 801763, 6 " Roller Pump Occlusion: Reference AA-2011-015 C. Indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

Recall: Z-1320-2012 · Initiated March 16, 2012

Recall

Recall Number
Z-1320-2012
Event Number
59573
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Device Design
Initiated
March 16, 2012
Posted
March 27, 2012
Terminated
May 2, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1 Base, 100/120V, Catalog 801763, 6 " Roller Pump Occlusion: Reference AA-2011-015 C. Indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

Reason

Terumo CVS has received reports of 27 instances in which the 6 inch Roller Pump jammed during cardioplegia delivery. There has been one report of the pump jamming during a case that concluded with an adverse patient outcome, although there was no indication that the pump jam contributed to the adverse patient outcome.

Action

"URGENT MEDICAL DEVICE RECALL" letters were sent on 3/16/12 to all consignees detailing the correction as well as also adding an Addendum to the Operator's manual for the Terumo System 1 to incorporate additional information and warnings for use when operating a 6" roller pump with dual-sized tubing. For questions call 1-800-521-2818.

Distribution

Worldwide Distribution--USA, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, Puerto Rico and the countries of INDIA, CANADA, CHILE, HONG KONG, EGYPT, PAKISTAN, KUWAIT, SAUDI ARABIA, JORDAN, UNITED ARAB EMIRATES, SOUTH AFRICA, SOUTH KOREA, PHILIPPINES, JAPAN, PHILIPPINES, SINGAPORE, AUSTRALIA, MALAYSIA, TAIWAN, RUSSIA, TURKEY, BELGIUM, GERMANY, COLUMBIA, COSTA RICA, ARGENTINA, HONDURAS, COLUMBIA, DOMINICAN REPUBLIC, GUATEMALA, MEXICO, VIETNAM, and THAILAND.

Quantity

994 units