FDA Recall Open, Classified

Spectral CT 7500 on Rails; Product Code (REF): 728334;

Recall: Z-1318-2026 · Initiated January 7, 2026

Recall

Recall Number
Z-1318-2026
Event Number
98248
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
JAK
Status
Open, Classified
Root Cause
Process change control
Initiated
January 7, 2026
Posted
February 6, 2026
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Spectral CT 7500 on Rails; Product Code (REF): 728334;

Reason

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Action

On or about January 7, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. As a precaution, monitor the gantry for any unusual sounds or vibrations during operation. If unusual sounds are heard or vibrations are observed, immediately call your service contact. In the absence of unusual sounds or vibrations, you may continue to use your system(s) according to the intended use. 2. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until Philips has inspected your system and has provided any correction, if needed. Ensure the letter is in a place likely to be seen and read. Actions planned by Philips: Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and to inspect the fasteners on your system (reference FCO72800845). If the inspection indicates that your system is affected, Philips will perform the correction to resolve the issue (reference FCO72800846)

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.

Quantity

3 units