FDA Recall Open, Classified

EPIX Therapeutics DiamondTemp Unidirectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF: CEDT200L, Sterile EO, UDI: 00812499030259

Recall: Z-1318-2021 · Initiated February 22, 2021

Recall

Recall Number
Z-1318-2021
Event Number
87412
Firm
EPIX THERAPEUTICS, INC
FEI Number
3015180993
Product Code
OAE
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 22, 2021
Address
945 Stewart Dr, Ste 100, Sunnyvale, CA, 94085-3940

Description

EPIX Therapeutics DiamondTemp Unidirectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF: CEDT200L, Sterile EO, UDI: 00812499030259

Reason

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

Action

On February 22, 2021, Medtronic issued a "Urgent Medical Device Correction" notification via regionally approved methods to consignees who have received affected product. In addition to informing consignees about the recall the firm asked consignees to take the following actions: 1. Review the corrected language applicable to IFU and label as provided in the Appendix below. Until the IFU is updated with the corrected information, the information provided in the Appendix should be used in combination with the existing labeling provided with the DiamondTemp" Catheters. 2. Please share this information with healthcare professionals in your facility that use DiamondTemp" Catheters. Also share this information with any other organization where these devices may have been transferred. 3. Please maintain a copy of this notice in your records. We are committed to patient safety and welcome any questions you may have regarding this communication. Please contact your Medtronic representative with questions.

Distribution

Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada

Quantity

60 units