FDA Recall Terminated

Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-800 series.

Recall: Z-1318-06 · Initiated June 17, 2005

Recall

Recall Number
Z-1318-06
Event Number
34953
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
June 17, 2005
Posted
August 2, 2006
Terminated
February 24, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-800 series.

Reason

Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.

Action

The customers were apprised of this situation with a recall letter issued by the firm on 6/17/2005. In addition, new Operators Manuals were sent to the customers. In the letter, the customers were advised that a Service Engineer would be contacting them shortly in order to arrange for the installation of the software upgrade. The firm''s Service Engineers visited each customer location and installed software upgrades. Version 3.831 software was installed on the 700 Series Analyzers and Version 4.14 was installed on the 800 Series Analyzers. According to the firm, these software corrections were verified as having been completed by 3/10/2006.

Distribution

Nationwide.

Quantity

108 devices