FDA Recall Terminated

ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..

Recall: Z-1316-2015 · Initiated February 16, 2015

Recall

Recall Number
Z-1316-2015
Event Number
70657
Firm
Brainlab AG Kapellenstrasse 12 Feldkirchen Germany
FEI Number
3002619595
Product Code
IYE
Status
Terminated
Root Cause
Software in the Use Environment
Initiated
February 16, 2015
Posted
March 25, 2015
Terminated
February 28, 2017

Description

ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..

Reason

The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.

Action

BrainLab sent an FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated February 16, 2015, to all affected customers. The letters included instructions for customers to: 1) Do not use any actively re-reconstructed CBCT subvolumes from the Varian TrueBeam system with the Brainlab ExacTrac 6.x CBCT Import & Alignment Software module; 2) Import exclusively original, not modified, CBCT volumes into ExacTrac to use for patient positioning; and, 3) Continue to verify any ExacTrac CBCT-based correction using the ExacTrac X-ray verification and/or an external IGRT procedure as mandatory. Brainlab provided existing potentially affected ExacTrac v.6.x CBCT Import & Alignment Software customers (with a Varian TrueBeam system) with the product notification information. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact the customer starting August 2015 to schedule the update installation. For further questions please call Customer Hotline + 49 89 99 15 68 44 or 1-800 597-5911 ( For US Customers only)

Distribution

Worldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.

Quantity

24 systems (US); 36 systems (Foreign)