FDA Recall Terminated

IntelliVue X3 Patient Monitor.

Recall: Z-1315-2018 · Initiated November 17, 2017

Recall

Recall Number
Z-1315-2018
Event Number
78732
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 17, 2017
Posted
April 6, 2018
Terminated
September 16, 2020
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

IntelliVue X3 Patient Monitor.

Reason

The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.

Action

Philips will contact customers with affected devices to arrange a correction of the issues. Firm plans a Software update of the affected devices and Replacement of the display assembly.

Distribution

48 Foreign Accounts

Quantity

975 (updated 8-14-18)