FDA Recall
Terminated
IntelliVue X3 Patient Monitor.
Recall: Z-1315-2018
·
Initiated November 17, 2017
Recall
- Recall Number
- Z-1315-2018
- Event Number
- 78732
- Firm
- Philips Electronics North America Corporation
- FEI Number
- 1218950
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- November 17, 2017
- Posted
- April 6, 2018
- Terminated
- September 16, 2020
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
IntelliVue X3 Patient Monitor.
Reason
The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.
Action
Philips will contact customers with affected devices to arrange a correction of the issues. Firm plans a Software update of the affected devices and Replacement of the display assembly.
Distribution
48 Foreign Accounts
Quantity
975 (updated 8-14-18)