FDA Recall Open, Classified

Abbott Point of Care Control Control i-STAT Level 1 10 X 1.7 mL

Recall: Z-1314-2024 · Initiated January 9, 2024

Recall

Recall Number
Z-1314-2024
Event Number
94024
Firm
Mckesson Medical-Surgical Inc. Corporate Office 9954 Maryland Drive Deep Run Iii Ste. 4000 Richmond VA 23233
FEI Number
3004386693
Product Code
JJY
Status
Open, Classified
Root Cause
Storage
Initiated
January 9, 2024
Posted
March 19, 2024

Description

Abbott Point of Care Control Control i-STAT Level 1 10 X 1.7 mL

Reason

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Action

McKesson notified customers by phone January 9, 2024 with a follow-up call January 17, 2024 instructing customers of three scenarios based upon the products they received and the investigation in conjunction with relevant manufacturer. The firm also sent Urgent Product Recall Notifications via email on February 8, 2024 instructing customers to immediately examine inventory and quarantine all affected devices, complete and return the provided response form to [email protected], and notify customers who you have further distributed recalled devices to of this recall and to segregate and quarantine any devices they have on hand. Replacement products will be provided and may have the same lot number(s) as the recalled items but are not affected by the recall and are acceptable for use. Customers with questions can contact McKesson directly by phone at (800) 688-8840. Representatives are available Monday to Friday, 8:00 am to 5:00 pm PST.

Distribution

US Nationwide distribution in the states of AZ, PA and TX.

Quantity

1 unit