FDA Recall Open, Classified

Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305

Recall: Z-1314-2023 · Initiated January 16, 2023

Recall

Recall Number
Z-1314-2023
Event Number
91557
Firm
Fujifilm Irvine Scientific, Inc.
FEI Number
2022379
Product Code
MQL
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
January 16, 2023
Posted
April 1, 2023
Address
2511 Daimler St, Santa Ana, CA, 92705-5533

Description

Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305

Reason

Due to the detection of oil toxicity in several lots of Oil for Embryo Culture products.

Action

On 01/16/2023, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, the firm has received complaints of reduced blastocyst development, including complete degradation in some cases, during culture of human embryos. Additionally, some clinics performed Human Sperm Survival Assay (HSSA), as part of incoming confirmatory Quality Control testing, and reported low sperm motility. Customers are instructed to: 1. Immediately stop using the product; 2. Quarantine any remaining product in inventory; 3. Return or destroy the product; 4. Complete and return the enclosed MEDICAL DEVICE RETURN RESPONSE FORM; and 5. Email the completed MEDICAL DEVICE RETURN RESPONSE FORM to [email protected] by January 23, 2023. For questions, contact: Customer Service Manager, Medical Business: - Toll free in the US: 1.800.577.6097 - Phone: 1.949.261.7800 x457 - Email: Melissa Gomez, [email protected]

Distribution

U.S.: AZ, AR, CA, CT, DC, FL, HI, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WI and WV. O.U.S.: India, Israel, Japan, Malaysia, Mexico, South Korea, South Africa, Thailand, Vietnam, Brazil, Bosnia, Argentina, Netherlands, and United Kingdom

Quantity

15,131 bottles