FDA Recall Terminated

D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.

Recall: Z-1314-06 · Initiated February 1, 2005

Recall

Recall Number
Z-1314-06
Event Number
35649
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CGZ
Status
Terminated
Root Cause
Other
Initiated
February 1, 2005
Posted
August 1, 2006
Terminated
November 18, 2011
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.

Reason

The CI membrane units generate too low values for status and sensitivity during calibration.

Action

The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.

Distribution

Nationwide.

Quantity

85/4 unit boxes