FDA Recall
Terminated
D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.
Recall: Z-1314-06
·
Initiated February 1, 2005
Recall
- Recall Number
- Z-1314-06
- Event Number
- 35649
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- CGZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 1, 2005
- Posted
- August 1, 2006
- Terminated
- November 18, 2011
- Address
- 810 Sharon Dr, Westlake, OH, 44145-1598
Description
D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.
Reason
The CI membrane units generate too low values for status and sensitivity during calibration.
Action
The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.
Distribution
Nationwide.
Quantity
85/4 unit boxes