FDA Recall
Terminated
medical device and software systems
Recall: Z-1314-04
·
Initiated June 24, 2004
Recall
- Recall Number
- Z-1314-04
- Event Number
- 29471
- Firm
- Varian Medical Systems Inc
- FEI Number
- 2916710
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 24, 2004
- Posted
- August 7, 2004
- Terminated
- December 23, 2004
- Address
- 3100 Hansen Way, Palo Alto, CA, 94304-1028
Description
medical device and software systems
Reason
a malfunction in the software of the VARiS versions 6.2.27 and 6.2.35 which is used to set up radiation therapy devices which will reset parameters to null for future treatments if a patient receives treatment on a non-varian machine and a varian machine and a non-varian field is loaded in the treatment session.
Action
Notification letter to consignees sent 06/14/2004.
Distribution
produt was distributed to 28 medical facilities, 12 in the United states, 16 foreign
Quantity
28 units