FDA Recall Terminated

Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200

Recall: Z-1313-06 · Initiated April 12, 2006

Recall

Recall Number
Z-1313-06
Event Number
35630
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
DHX
Status
Terminated
Root Cause
Other
Initiated
April 12, 2006
Posted
August 1, 2006
Terminated
April 3, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200

Reason

Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens.

Action

A Product Corrective Action letter was sent via US mail the week of April 10, 2006 to all Access CEA customers informing them that the Access CEA Assay could generate decreased values with certain frozen patient specimens. Although they have found it at a low occurrence, it is recommended that specimens should not be frozen prior to analysis until further notice. The letter informs customers that Beckman Coulter has verified sample stability through at least 14 days when properly stored at 2-8 degrees Celsius and shipping stability for 4 days at 2-8 degrees Celsius using cool packs. Firm also advises customers to contact customer technical support or local Beckman Coulter Representative if laboratory is unable to avoid freezing specimens and to review results from frozen CEA samples to ensure that the match the total clinical presentation per the customer''s established laboratory procedure. The letter requests that the notification remain part of the laboratory Quality System documentation and that the response form be returned to the firm within 10 days.

Distribution

Nationwide and Canada

Quantity

775