FDA Recall Open, Classified

Numerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM; (2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM; (3) REF BH957R, Schnidt Tonsil Forceps SLT-CVD185MM; (4) REF BH959R, Schnidt Tonsil Forceps CVD 185MM; (5) REF BH961R, Schnidt Tonsil FCPSOPEN-RINGCVD 185MM; (6) REF BH963R, Schnidt Tonsil FCPSOPEN-RINGCVD185MM; and (7) REF BH965R, Wilson Tonsil Forceps 190MM.

Recall: Z-1312-2025 · Initiated January 31, 2025

Recall

Recall Number
Z-1312-2025
Event Number
96223
Firm
Aesculap Inc
FEI Number
2916714
Product Code
KAE
Status
Open, Classified
Root Cause
Use error
Initiated
January 31, 2025
Posted
March 11, 2025
Address
3773 Corporate Pkwy, Center Valley, PA, 18034-8217

Description

Numerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM; (2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM; (3) REF BH957R, Schnidt Tonsil Forceps SLT-CVD185MM; (4) REF BH959R, Schnidt Tonsil Forceps CVD 185MM; (5) REF BH961R, Schnidt Tonsil FCPSOPEN-RINGCVD 185MM; (6) REF BH963R, Schnidt Tonsil FCPSOPEN-RINGCVD185MM; and (7) REF BH965R, Wilson Tonsil Forceps 190MM.

Reason

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Action

The recalling firm issued letters dated 1/29/2025 on 1/31/2025 via certified mail return-receipt. The letter explained the reason for the correction, identified the product impacted, provided the risk to health, and the actions required for the customer/user. The actions included: (1) Review the Urgent Medical Device Correction Notification in its entirety and ensure all users within the consignee organization are informed about the correction. Post the notification where the affected products are stored. If the consignee is a distributor, they are to forward the notification to their customers. The correction is to be extended to the end user/customer level; (2) Determine the current inventory of the affected items and comply with the following instructions: (a) Use the products identified for clamping of tissue and small vessels only; (b) Inspect the products for end-of-life indicators prior to use; (c) Visually inspect the products which should be clean and free of any signs of corrosion and cracks. Check the joint and jaw in the open position from both sides (if needed, use magnification; (d) Test the function of the device, including the jaw area, for deformation and loss of tension. An Acknowledgment Form was enclosed for completion and return via FAX or email within 2 weeks of receipt, even if the total inventory in the consignee location is 0.

Distribution

Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.

Quantity

47,078 forceps