FDA Recall Open, Classified

Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,

Recall: Z-1310-2023 · Initiated February 24, 2023

Recall

Recall Number
Z-1310-2023
Event Number
91846
Firm
Iradimed Corporation
FEI Number
3005053560
Product Code
FRN
Status
Open, Classified
Root Cause
Process control
Initiated
February 24, 2023
Posted
March 30, 2023
Address
1025 Willa Springs Dr, Winter Springs, FL, 32708-5235

Description

Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,

Reason

Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.

Action

Iradimed Corporation issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/24/2023 by certified mail. The notice explained the issue and the risk and provided updated instructions for use to mitigate the potential risk of inlet occlusion. For further information concerning this issue or have any questions, please contact Iradimed support via email at [email protected] or via telephone at 407-677-8022.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates.

Quantity

3285 boxes of 50 devices