FDA Recall Terminated

MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated

Recall: Z-1309-2016 · Initiated March 7, 2016

Recall

Recall Number
Z-1309-2016
Event Number
73550
Firm
Viewray Incorporated
FEI Number
3007546534
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
March 7, 2016
Posted
March 31, 2016
Terminated
February 22, 2017
Address
2 Thermo Fisher Way, Oakwood Village, OH, 44146-6536

Description

MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated

Reason

When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location.

Action

ViewRay sent an Urgent Device Correction letter dated March 7, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers with questions were instructed to contact ViewRay Customer Support at [email protected] or call the ViewRay Customer Support center at 855-286-8875.

Distribution

Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.

Quantity

5 units