MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated
Recall
- Recall Number
- Z-1309-2016
- Event Number
- 73550
- Firm
- Viewray Incorporated
- FEI Number
- 3007546534
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 7, 2016
- Posted
- March 31, 2016
- Terminated
- February 22, 2017
- Address
- 2 Thermo Fisher Way, Oakwood Village, OH, 44146-6536
Description
MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated
When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location.
ViewRay sent an Urgent Device Correction letter dated March 7, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers with questions were instructed to contact ViewRay Customer Support at [email protected] or call the ViewRay Customer Support center at 855-286-8875.
Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.
5 units