FDA Recall Open, Classified

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Recall: Z-1308-2026 · Initiated November 17, 2025

Recall

Recall Number
Z-1308-2026
Event Number
98087
Firm
ETAC A/S Parallelvej 3 Gedved Denmark
FEI Number
3004137175
Product Code
FSA
Status
Open, Classified
Root Cause
Component design/selection
Initiated
November 17, 2025
Posted
February 5, 2026

Description

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Reason

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Action

On November 17, 2025 customers were notified vis a phone call. Notifying them that they will receive an email with the Field Safety Notice and Response Form. The customers were emailed Field Safety Notice letters on or around November 20, 2025. Corrective action: To reduce the risk of further incidents, replacement units are offered for all potentially affected products. Continued use is safe as long as the lifting bar can move freely and remains vertical during the lifting cycle with or without load. If this is not the case, or if there is any doubt about the product s functionality, the unit must be taken out of service immediately. The components in the defective units (Mover 205 and Mover 300) will gradually be replaced in the coming months. The affected lifting bars will likewise be replaced.

Distribution

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Quantity

95 units