FDA Recall Open, Classified

Numerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM; (2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM; (3) REF BJ511R, Holzbach Hyst Forceps CVD245MM; (4) REF BJ520R, Heaney Hyst Forceps Grooved 205MM; (5) REF BJ521R, Heaney Hyst Forceps Grooved 250MM; (6) REF BJ522R, Heaney Hyst Forceps Grooved 195MM; (7) REF BJ524R, Heaney-Rezek Fcpsstr Ser Lgth 210MM; (8) REF BJ527R, Wiener Hyst Forceps Grooves 240MM; (9) REF BJ530R, Heaney Hysterectomy Forceps STR 225MM; (10) REF BJ531R, Heaney Hysterectomy Forceps CVD200MM; (11) REF BJ532R, Heaney Hysterectomy Forceps CVD215MM; (12) REF BJ534R, Hysterectomy For.Gwilliamsstr.205MM; (13) REF BJ550R, Wertheim Parametrium Clamp STR 250MM; (14) REF BJ551R, Wertheim Parametrium Clamp CVD245MM; (15) REF BJ552R, Wertheim Parametrium Clamp CVD240MM; (16) REF BJ553R, Wertheim Parametrium Clamphvy-CVD220MM; (17) REF MD591, Heany Ballentine Hyst Fcps STR.215MM; and (18) REF MD592, Heany Ballentine Hyst Fcps CVD.215MM.

Recall: Z-1308-2025 · Initiated January 31, 2025

Recall

Recall Number
Z-1308-2025
Event Number
96223
Firm
Aesculap Inc
FEI Number
2916714
Product Code
HGC
Status
Open, Classified
Root Cause
Use error
Initiated
January 31, 2025
Posted
March 11, 2025
Address
3773 Corporate Pkwy, Center Valley, PA, 18034-8217

Description

Numerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM; (2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM; (3) REF BJ511R, Holzbach Hyst Forceps CVD245MM; (4) REF BJ520R, Heaney Hyst Forceps Grooved 205MM; (5) REF BJ521R, Heaney Hyst Forceps Grooved 250MM; (6) REF BJ522R, Heaney Hyst Forceps Grooved 195MM; (7) REF BJ524R, Heaney-Rezek Fcpsstr Ser Lgth 210MM; (8) REF BJ527R, Wiener Hyst Forceps Grooves 240MM; (9) REF BJ530R, Heaney Hysterectomy Forceps STR 225MM; (10) REF BJ531R, Heaney Hysterectomy Forceps CVD200MM; (11) REF BJ532R, Heaney Hysterectomy Forceps CVD215MM; (12) REF BJ534R, Hysterectomy For.Gwilliamsstr.205MM; (13) REF BJ550R, Wertheim Parametrium Clamp STR 250MM; (14) REF BJ551R, Wertheim Parametrium Clamp CVD245MM; (15) REF BJ552R, Wertheim Parametrium Clamp CVD240MM; (16) REF BJ553R, Wertheim Parametrium Clamphvy-CVD220MM; (17) REF MD591, Heany Ballentine Hyst Fcps STR.215MM; and (18) REF MD592, Heany Ballentine Hyst Fcps CVD.215MM.

Reason

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Action

The recalling firm issued letters dated 1/29/2025 on 1/31/2025 via certified mail return-receipt. The letter explained the reason for the correction, identified the product impacted, provided the risk to health, and the actions required for the customer/user. The actions included: (1) Review the Urgent Medical Device Correction Notification in its entirety and ensure all users within the consignee organization are informed about the correction. Post the notification where the affected products are stored. If the consignee is a distributor, they are to forward the notification to their customers. The correction is to be extended to the end user/customer level; (2) Determine the current inventory of the affected items and comply with the following instructions: (a) Use the products identified for clamping of tissue and small vessels only; (b) Inspect the products for end-of-life indicators prior to use; (c) Visually inspect the products which should be clean and free of any signs of corrosion and cracks. Check the joint and jaw in the open position from both sides (if needed, use magnification; (d) Test the function of the device, including the jaw area, for deformation and loss of tension. An Acknowledgment Form was enclosed for completion and return via FAX or email within 2 weeks of receipt, even if the total inventory in the consignee location is 0.

Distribution

Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.

Quantity

19,979 forceps