FDA Recall Open, Classified

VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961 VLT600DF AIM STP ARDVST229000A VST60SF AIM ARDVST229001A VST60DF AIM ARDVST229002A VST66SF AIM ARDVST229003A VST66DF AIM ARDVST229034A VST60SF ARDVST229036A VST66SF ARDVST229037A VST66DF

Recall: Z-1308-2022 · Initiated May 6, 2022

Recall

Recall Number
Z-1308-2022
Event Number
90305
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
FSY
Status
Open, Classified
Root Cause
Process change control
Initiated
May 6, 2022
Posted
June 30, 2022
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961 VLT600DF AIM STP ARDVST229000A VST60SF AIM ARDVST229001A VST60DF AIM ARDVST229002A VST66SF AIM ARDVST229003A VST66DF AIM ARDVST229034A VST60SF ARDVST229036A VST66SF ARDVST229037A VST66DF

Reason

Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the patient body or in the operating field.

Action

On May 6, 2022, the firm notified affected customers through an Urgent Medical Device Correction letter. Customers were instructed to continue to perform daily inspections per the IFU, which include checking for any chipped or missing paint. If you observe paint chipping or detachment, the firm recommends that you stop using the device and contact your Getinge service representative to schedule an appointment to replace the part free of charge. The firm is requesting that customers contact their Getinge service representative to have the part replaced if chipped or missing paint is observed through April 2025. Getinge service representatives will inspect the affected component during periodic maintenance and will replace the component if the issue is observed through April 2025.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, GA, IA, ID, KS, KY, LA, MA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, VA, WA.

Quantity

337 (US); 8594 (OUS)