VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Recall
- Recall Number
- Z-1308-2016
- Event Number
- 73436
- Firm
- MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK 6-22, 4-CHOME, KAN-ON-SHIN-MACHI NISHI-KU HIROSHIMA Japan
- FEI Number
- 3006942329
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 7, 2016
- Posted
- March 31, 2016
- Terminated
- April 20, 2017
Description
VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.
A Field Safety Notice dated March 2, 2016, was sent to all affected customers via e-mail on March 7, 2016. The Field Safety Notice informed users of the problem and provided a safe workaround until corrected software is available. The correction will be free of charge and a Customer Service Representative will contact all affected customers to schedule its installation. Customers with questions were advised to contact their Customer Service Representative.
Nationwide Distribution including New York, Texas, Florida, and Ohio
25 Worldwide, US: 4 units