FDA Recall Terminated

VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Recall: Z-1307-2021 · Initiated January 28, 2021

Recall

Recall Number
Z-1307-2021
Event Number
87249
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
LON
Status
Terminated
Root Cause
Software design
Initiated
January 28, 2021
Terminated
January 12, 2023
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Reason

bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

Action

The customer letter instructs the customer on how to verify the external communication configuration setting to see if they are impacted by this correction. If impacted, the customer will need to contact their local bioMerieux representative for instructions on how to verify expertised results until a correction is implemented. For a correction, customers can either convert to MYLA or BCI (where this issue will not occur) or continue to verify the expertised results until a software patch is deployed for 9.0x customers.

Distribution

Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.

Quantity

8786 systems