FDA Recall Terminated

Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065 and 9067. These rechargeable batteries are used to supply power to the following products: Wireless ECG Module used with the Precess MRI compatible Patient Monitoring System, Wireless SpO2 Module used with the Precess MRI Compatible Patient Monitoring System, Philips Basic Triggering Set, Philips Monitoring Sensor Set, Philips Battery Set

Recall: Z-1307-2015 · Initiated November 20, 2014

Recall

Recall Number
Z-1307-2015
Event Number
69812
Firm
Philips Medical Systems, Inc.
FEI Number
1218950
Product Code
LNH
Status
Terminated
Root Cause
Component design/selection
Initiated
November 20, 2014
Posted
March 20, 2015
Terminated
January 23, 2017
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065 and 9067. These rechargeable batteries are used to supply power to the following products: Wireless ECG Module used with the Precess MRI compatible Patient Monitoring System, Wireless SpO2 Module used with the Precess MRI Compatible Patient Monitoring System, Philips Basic Triggering Set, Philips Monitoring Sensor Set, Philips Battery Set

Reason

VCG battery ignited in a VCG unit when recharging.

Action

Consignees were informed of the recall via letter sent on November 20, 2014 via certified mailing. Field Service Engineers (FSE) will replace all affected batteries.

Distribution

Distributed in the following states: AL, CA, FL, GA, IL, MA, MN, NC, ND, NE, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI; OUS: Australia, Austria, Belgium, China, Ecuador, Finland, France, Germany, Italy, Japan, Mexico, Netherlands, slovenia, South Africa, Sweden, Switzerland, United Kingdom

Quantity

79