FDA Recall Terminated

RayStation Radiation Therapy Treatment Planning System, models 2.5, 3.0, 3.5 and 4.0; a software system designed for treatment planning and analysis of radiation therapy.

Recall: Z-1307-2014 · Initiated February 6, 2014

Recall

Recall Number
Z-1307-2014
Event Number
67504
Firm
RAYSEARCH LABORATORIES AB Sveavaegen 9 Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Terminated
Root Cause
Software Design Change
Initiated
February 6, 2014
Posted
March 28, 2014
Terminated
February 17, 2017

Description

RayStation Radiation Therapy Treatment Planning System, models 2.5, 3.0, 3.5 and 4.0; a software system designed for treatment planning and analysis of radiation therapy.

Reason

Potential for dose errors due to software program errors.

Action

RaySearch Laboratories sent an Field Safety Notice, Medical Device Correction letter dated February 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. ACTIONS TO BE TAKEN BY THE USER A workaround that eliminates the described error is to select the discard rotations option in the Image Registration Module before creating a rigid registration intended to be used for hybrid deformable registration. Another option is to use the ROI based deformable registration when applicable, as this functionality is not affected by the error. SOLUTION This issue will be resolved in a patch release of RayStation, scheduled for market release February 28, 2014. Release schedule is subject to regulatory approval for some markets. In the meantime, this field safety notice is being distributed to all customers. CONTACT INFORMATION For further information, please contact For the Americas: Customer Support, RaySearch Americas, at +1 877 778 3849 For Europe, Asia and rest of the world: RaySearch Laboratories AB, at +46 8 5450 6130 Thank you for your cooperation, and we apologize for any inconvenience

Distribution

Worldwide Distribution - US including the states of NY, OH, MD, CT, CO, TN, IL, AZ, MS, NJ, PA, DE, FL, CA and WA., and the countries of New Zealand, Australia, Republic of Korea, China, Italy, Germany, Norway, Belgium, Switzerland, United Kingdom and Spain.

Quantity

151 units Total (88 units domestically & 63 units internationally)