FDA Recall Open, Classified

NaviCare Nurse Call/Voalte Nurse Call

Recall: Z-1306-2022 · Initiated May 31, 2022

Recall

Recall Number
Z-1306-2022
Event Number
90322
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
ILQ
Status
Open, Classified
Root Cause
Software Design Change
Initiated
May 31, 2022
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

NaviCare Nurse Call/Voalte Nurse Call

Reason

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Action

The firm disseminated an Urgent Medical Device Correction Notice beginning on 05/31/2022 by letter. The letter described the problem and requested the following actions be taken by the customer: "Identify if your Nurse Call System utilizes one of the impacted wireless integrations identified. If so, please share the information with potential users in your organization. Hillrom recommends reviewing the wireless system settings with your wireless provider to determine if there are configurable options for call acceptance only. Complete the enclosed response form and return to [email protected] within two weeks of receipt of this notice, to confirm you have received this communication and understand the potential issue." Persons with questions were directed as follows: "If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support, available twenty-four hours a day, seven days a week, at 1-800-445-3720, option 3, or [email protected]"

Distribution

nationwide and Canada

Quantity

283 installations