FDA Recall Terminated

Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics International, Long Jing Industrial Estate, Long Jing Road, Xi Li Town, Nantou District, Shenzhen, China

Recall: Z-1306-06 · Initiated May 22, 2006

Recall

Recall Number
Z-1306-06
Event Number
35486
Firm
Abbott Diabetes Care, Inc.
FEI Number
3002803444
Product Code
NBW
Status
Terminated
Root Cause
Other
Initiated
May 22, 2006
Posted
August 3, 2006
Terminated
December 21, 2006
Address
1360 S Loop Rd, Alameda, CA, 94502-7000

Description

Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502. Manufactured by: Flextronics International, Long Jing Industrial Estate, Long Jing Road, Xi Li Town, Nantou District, Shenzhen, China

Reason

The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration data including selectable unit of measure, and strip calibration code.

Action

On 5/22/06, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions for the corrective actions.

Distribution

Worldwide

Quantity

approx. 2,365,000 meters (both recall numbers)