FDA Recall
Terminated
TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach.
Recall: Z-1305-06
·
Initiated April 6, 2006
Recall
- Recall Number
- Z-1305-06
- Event Number
- 35280
- Firm
- Kensey Nash Corp
- FEI Number
- 1000307073
- Product Code
- NFA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 6, 2006
- Posted
- July 28, 2006
- Terminated
- August 10, 2006
- Address
- 735 Pennsylvania Dr, Exton, PA, 19341-1130
Description
TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach.
Reason
Alarm activiation-A priming issue involving the flow control unit due to a software problem has caused false postive Extraction Line Block (ELB) alarming. Use of this product may result in a health hazard resulting from early termination of the flush and extraction procedure after stenting
Action
The recalling firm issued a draft recall letter to their European distributor who translated the letter for their customers.
Distribution
The product was shipped to hospitals in Germany and Italy.
Quantity
69 kits