FDA Recall Terminated

TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach.

Recall: Z-1305-06 · Initiated April 6, 2006

Recall

Recall Number
Z-1305-06
Event Number
35280
Firm
Kensey Nash Corp
FEI Number
1000307073
Product Code
NFA
Status
Terminated
Root Cause
Other
Initiated
April 6, 2006
Posted
July 28, 2006
Terminated
August 10, 2006
Address
735 Pennsylvania Dr, Exton, PA, 19341-1130

Description

TriActiv ProGuard Embolic Protection System, Part number 61000-01. The product is shipped as a procedure kit. It is used with a standard 6F guide catheter using the standard femoral approach.

Reason

Alarm activiation-A priming issue involving the flow control unit due to a software problem has caused false postive Extraction Line Block (ELB) alarming. Use of this product may result in a health hazard resulting from early termination of the flush and extraction procedure after stenting

Action

The recalling firm issued a draft recall letter to their European distributor who translated the letter for their customers.

Distribution

The product was shipped to hospitals in Germany and Italy.

Quantity

69 kits