FDA Recall Terminated

Dynavox T10 Speech Generating Device ( including Batteries)

Recall: Z-1304-2018 · Initiated December 14, 2017

Recall

Recall Number
Z-1304-2018
Event Number
79433
Firm
Dynavox Systems Llc
FEI Number
3004959902
Product Code
ILQ
Status
Terminated
Root Cause
Component design/selection
Initiated
December 14, 2017
Terminated
December 11, 2019
Address
2100 Wharton St, Ste Ste 400, Pittsburgh, PA, 15203-1691

Description

Dynavox T10 Speech Generating Device ( including Batteries)

Reason

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Action

On February 26, 2018 an URGENT MEDICAL RECALL LETTER was issued to customers with the subject line listing the serial number specific to that customer along with the date of purchase. The letter informs customers of the issue and instructs customers to return product to one of their facilities to have the product repaired until an RA. Questions or concerns can be directed to 1-800-344-1778, Press 1 and then Press 2.

Distribution

Nationally

Quantity

4615