FDA Recall Open, Classified

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Recall: Z-1303-2024 · Initiated February 8, 2024

Recall

Recall Number
Z-1303-2024
Event Number
94012
Firm
Beckman Coulter, Inc.
FEI Number
2122870
Product Code
KLI
Status
Open, Classified
Root Cause
Vendor change control
Initiated
February 8, 2024
Posted
March 15, 2024
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037

Description

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Reason

Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

Action

The recalling firm issued letters on 2/8/2024 via email and standard postal mail to their U.S. customers and postal mail to their OUS customers. The letter explained the issue, impact, and actions to be taken. The actions included: (1) Discontinue using the affected lots and discard all remaining reagent packs from the lots; (2) The firm recommended performing a retrospective review on patient results generated with the affected reagent lots and assess the need for retesting;(3) Contact the local Beckman Coulter representative for replacement product requests; (4) For customers in the U.S. and Canada, complete the attached Replacement Order Form; and (5) For customers outside the U.S. and Canada, contact their local Beckman Coulter representative. The consignee was requested to share the letter information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If the affected product was forwarded to another laboratory, they were requested to provide them a copy of the letter. The consignee was requested to respond within 10 days either electronically via email or manually by completing and returning the enclosed Response Form.

Distribution

Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.

Quantity

64,009 kits