Syngo.via. Medical Device Software. Picture archiving and communications system.
Recall: Z-1303-2018
·
Initiated January 8, 2018
Recall
- Recall Number
- Z-1303-2018
- Event Number
- 79532
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 8, 2018
- Posted
- March 20, 2018
- Terminated
- April 22, 2020
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Syngo.via. Medical Device Software. Picture archiving and communications system.
Reason
A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.
Action
Customer Advisory Notice SY065/17/P was sent to direct accounts. Firm is continuing to work on software solution. A workaround is available and is described in the Customer Advisory Notice.
Distribution
Device is software only. No products are distributed to wholesale dealers, distributers or retailers
Quantity
236 units