FDA Recall Open, Classified

Volcano Visions PV. 035 Digital IVUS Catheter REF 81234, PN: 300007367341 , Rx Only, Sterile EO, CE 2797

Recall: Z-1302-2022 · Initiated May 13, 2022

Recall

Recall Number
Z-1302-2022
Event Number
90252
Firm
Volcano Corp
FEI Number
3008363989
Product Code
OBJ
Status
Open, Classified
Root Cause
Process control
Initiated
May 13, 2022
Address
3721 Valley Centre Dr, Ste 500, San Diego, CA, 92130-3328

Description

Volcano Visions PV. 035 Digital IVUS Catheter REF 81234, PN: 300007367341 , Rx Only, Sterile EO, CE 2797

Reason

Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications

Action

On May 25, 2022, Philips issued a "Urgent Medical Device Correction" notification. Phillips asked customers to take the following actions: 1. Philips recommends reviewing the enclosed compliant Instructions for Use, in Appendix A, before proceeding as normal. 2. As per standard practice, precaution should always be used when advancing or removing a catheter in complex vessel anatomies, such as not forcing a catheter into a narrow vessel or a tight stenosis. Vessel calcification, tortuosity and pre-existing untreated vessel spasm are key components to any complex vascular anatomy. Additionally, if the initial guidewire faces resistance or requires further manipulations, please use this as an indicator of complex patient anatomy. 3. Philips is providing this Urgent Medical Device Correction letter to affected customers and recommends forwarding this notification to all staff who need to be aware within your organization or to any organization where the potentially affected devices may have been transferred. 4. To acknowledge receipt of this notification, please complete, sign, and return the Customer Reply Form within 30 days upon receipt of this notice to Email: [email protected]

Distribution

Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, FR, PR, IT, DE, CA, NZ, PL, BR, ZA, NL.

Quantity

7,520 units