FDA Recall Terminated

GE Healthcare Centricity PACS-IW, Model Numbers: (a) 2052831-00X (b) 2049588-008 Product Usage: Centricity PACS-IW by GE Healthcare is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Recall: Z-1302-2018 · Initiated February 23, 2018

Recall

Recall Number
Z-1302-2018
Event Number
79537
Firm
GE Healthcare
FEI Number
3004526608
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
February 23, 2018
Terminated
September 30, 2024
Address
540 W Northwest Hwy, Barrington, IL, 60010-3051

Description

GE Healthcare Centricity PACS-IW, Model Numbers: (a) 2052831-00X (b) 2049588-008 Product Usage: Centricity PACS-IW by GE Healthcare is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Reason

A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more image series (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series.

Action

GE Healthcare sent an Urgent Medical Device Correction letter dated February 23, 2018 to their customers. The letter provided the affected product, problem and provided the following safety instructions: "Until a product modification is available to correct this issue, the following actions are recommended: 1. Utilize the image count within the QC process to alert the user of a discrepancy in the number of transmitted images from the modality to the number of images available in the Universal Viewer. a. If a discrepancy is identified, attempt to retransmit the exam to PACS. b. If retransmission is unsuccessful, a GE Healthcare Service representative will need to be contacted to assist with resolution of the impacted exam. Urgent cases impacted by this issue should be interpreted at the modality. 2. If DICOM storage commit is configured and in use, no commit notification is sent to the modality for images that were impacted by this issue. 3. Attention should be given to prior cases, acquired before the application of the safety instructions given in this communication as they may be affected by the safety issue." If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution-US Nationwide

Quantity

569 units