FDA Recall Terminated

da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.

Recall: Z-1302-2015 · Initiated March 3, 2015

Recall

Recall Number
Z-1302-2015
Event Number
70705
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 3, 2015
Posted
March 20, 2015
Terminated
September 8, 2015
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304

Description

da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.

Reason

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

Action

The firm, Intuitive Surgical, sent an "Urgent Medical Device Correction" letter dated March 11, 2015 to its consignees/customers on March 11, 2015 via Fed Ex. The letter described the product, problem and actions to be taken. The customers were instructed to ensure that all affected personnel are fully informed of this notice; forward this notice to your Risk Manager, OR Director, Purchasing Manager, biomedical Engineering staff and members of your medical staff who perform da Vinci Surgery procedures; Identify systems containing affected product at your site; if you encounter this error during a procedure and the fault cannot be recovered, disable the affected arm and contact dVStat; Contact Intuitive Surgical Customer Service; Complete and return the attached Acknowledgment Form to Intuitive Surgical via U.S. Fax +1 (408) 716-3040, or Scan and Email: [email protected], ATTN: REGULATORY COMPLIANCE-Subject line for email: Pot to Encoder;and retain a copy of this notice for your records. An Intuitive Surgical representative will contact you to schedule a remote assessment to determine the health of your system and to schedule an appointment to replace the affected arm. If you need further information or support concerning this Medical Device Correction, please contact your Clinical Sales Representative or Intuitive Surgical Customer Service at the numbers listed below: -North America and South America: 800-876-1310, Option 3 (6 AM to 5 PM PST) -Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) -South Korea: 02-3271-3200 (9 AM to 6 PM KST) -Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or [email protected].

Distribution

Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Quantity

167 total da Vinci Xi affected systems