FDA Recall Terminated

BioMend Absorbable Collagen Membrane BioMend Absorbable Collagen Membrane is an absorbable, implantable material that is indicated for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus.

Recall: Z-1301-2013 · Initiated March 28, 2013

Recall

Recall Number
Z-1301-2013
Event Number
64778
Firm
Integra LifeSciences Corporation
FEI Number
1121308
Product Code
LYC
Status
Terminated
Root Cause
Process control
Initiated
March 28, 2013
Posted
May 10, 2013
Terminated
June 3, 2015
Address
105 Morgan Ln, Plainsboro, NJ, 08536-3339

Description

BioMend Absorbable Collagen Membrane BioMend Absorbable Collagen Membrane is an absorbable, implantable material that is indicated for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus.

Reason

Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.

Action

The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: [email protected] or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol." For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723.

Distribution

Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.

Quantity

32,446 boxes