FDA Recall
Terminated
ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Spatula; Cat. #60-5158-234 -''J'' Hook; Cat. #60-5158-432 - ''L'' Hook; Cat. # 60-5158-934 - Needle. FIRM ON LABEL: ConMed Corporation, Utica, New York, USA 13502.
Recall: Z-1301-04
·
Initiated June 28, 2004
Recall
- Recall Number
- Z-1301-04
- Event Number
- 29478
- Firm
- Conmed Corporation
- FEI Number
- 3001451571
- Product Code
- GEI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 28, 2004
- Posted
- July 31, 2004
- Terminated
- October 5, 2005
- Address
- 525 French Rd, Utica, NY, 13502-5945
Description
ConMed Stealth coated Laparoscopic Electrodes with 4mm Pin (60-5158 series). Cat. #60-5158-034 - Spatula; Cat. #60-5158-234 -''J'' Hook; Cat. #60-5158-432 - ''L'' Hook; Cat. # 60-5158-934 - Needle. FIRM ON LABEL: ConMed Corporation, Utica, New York, USA 13502.
Reason
Insulation of some electrodes may have pulled away from the hub of the electrode exposing the conductive shaft below. There is potential for electrical shock.
Action
Letters dated 6/28/04 via Priority Mail with instructions to return product and to subrecall if further distributed.
Distribution
Nationwide distribution to 24 hospitals, 3 distributors and 13 sales reps. Six foreign consignees in Mexico, Italy, United Kingdom, Spain, Turkey and Australia.
Quantity
3,252 units