FDA Recall Open, Classified

GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed

Recall: Z-1299-2025 · Initiated February 18, 2025

Recall

Recall Number
Z-1299-2025
Event Number
96370
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
JAK
Status
Open, Classified
Root Cause
Process change control
Initiated
February 18, 2025
Posted
March 7, 2025
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed

Reason

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to [email protected]."

Distribution

Worldwide distribution.

Quantity

42 units