SPII Model Lubinus, Long Stem Prosthesis Standard Neck, cemented

Recall: Z-1298-2024 · Initiated February 6, 2024

Recall

Recall Number
Z-1298-2024
Event Number
94078
Firm
Waldemar Link GmbH & Co. KG (Mfg Site) Oststr. 4-10 Norderstedt Germany
FEI Number
3003386935
Product Code
JDG
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 6, 2024
Posted
March 14, 2024

Description

SPII Model Lubinus, Long Stem Prosthesis Standard Neck, cemented

Reason

Due to two complaints, has come to our attention that additional guidance is required on the correct interpretation of the carton label Size and Type columns to avoid misinterpretation during surgery.

Action

Urgent Medical Device Correction (UMDC) Letters were sent via email to consignees on February 16, 2024. Actions to be taken by the customer/user: Please take note of the clarification of the carton label noted in section 2. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to correctly interpret the Size and Type columns on the label. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, NJ, NV, PA, TN, TX, VA, WA.

Quantity

153 Units