Operating Table System

Recall: Z-1296-2021 · Initiated February 11, 2021

Recall

Recall Number
Z-1296-2021
Event Number
87386
Firm
Getinge Group Logistics America, LLC
FEI Number
3025504317
Product Code
FQO
Status
Open, Classified
Root Cause
Device Design
Initiated
February 11, 2021
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Operating Table System

Reason

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Action

In February, 2021, Maquet GmbH/Getinge, will issue an Urgent Medical Device Correction notice to customers via FedEx 2 Day Delivery for the Magnus 1180 Operating Table due to a potential for hydraulic leakage to occur at the inclination function which may result in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. If this were to occur, it may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Distribution

Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.

Quantity

4288 units