OraTape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings for Dental Surgery are intended for use on moist or bleeding oral wounds where a stabilized blood clot can help protect the surface of the wound from further injury. Topical administration of the Absorbable Collagen Wound Dressings directly over the wound effectively controls bleeding usually within two to five minutes.
Recall
- Recall Number
- Z-1296-2013
- Event Number
- 64778
- Firm
- Integra LifeSciences Corporation
- FEI Number
- 1121308
- Product Code
- LPG
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 28, 2013
- Posted
- May 10, 2013
- Terminated
- June 3, 2015
- Address
- 105 Morgan Ln, Plainsboro, NJ, 08536-3339
Description
OraTape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings for Dental Surgery are intended for use on moist or bleeding oral wounds where a stabilized blood clot can help protect the surface of the wound from further injury. Topical administration of the Absorbable Collagen Wound Dressings directly over the wound effectively controls bleeding usually within two to five minutes.
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: [email protected] or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol." For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723.
Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.
17,110 boxes