FDA Recall Open, Classified

Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Recall: Z-1294-2018 · Initiated January 1, 2017

Recall

Recall Number
Z-1294-2018
Event Number
79362
Firm
Seiler Instr. & Mfg. Co. Inc.
FEI Number
1927430
Product Code
EPT
Status
Open, Classified
Root Cause
Use error
Initiated
January 1, 2017
Address
3433 Tree Court Industrial Blvd, Saint Louis, MO, 63122-6617

Description

Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Reason

The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.

Action

The customer was notified via email on about January 4, 2017. Instructions included that the scope only tips if the user has it extended all the way out and is pulling on the handles of the scope. The end user must not pull on the handles of the scope when it is extended. They were also instructed to educate personnel on this matter. For further questions, please call (314) 968-2282.

Distribution

Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.

Quantity

43 units, but 1 affected