FDA Recall Terminated

Sarns Soft-Flow Aortic Cannula without suture flange Angled tip, wire-reinforced with luer port 8.0mm (24Fr) OD with 3/8" connector, 14" (36cm) Product Usage: CTS Usage: The Sarns-Aortic Cannula is indicated for use In perfusion of the ascending aorta during cardiopulmonary bypass surgery. The X-Coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures. long, with Xcoating" surface coating

Recall: Z-1294-2012 · Initiated February 23, 2012

Recall

Recall Number
Z-1294-2012
Event Number
60718
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DWF
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 23, 2012
Posted
March 23, 2012
Terminated
April 3, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Sarns Soft-Flow Aortic Cannula without suture flange Angled tip, wire-reinforced with luer port 8.0mm (24Fr) OD with 3/8" connector, 14" (36cm) Product Usage: CTS Usage: The Sarns-Aortic Cannula is indicated for use In perfusion of the ascending aorta during cardiopulmonary bypass surgery. The X-Coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures. long, with Xcoating" surface coating

Reason

During in-house inspection, Terumo Cardiovascular Systems discovered one Soft-Flow Aortic Cannula with a hole in the tip, just above the suture ball.

Action

Terumo sent an URGENT MEDICAL DEVICE RECALL letter dated February 23, 2012 via Federal Express to all US customers. Terumo Foreign Affiliates received the recall notification via email on February 24th, 2012. The letters identified the affected products, reson for removal, potential hazard, affected population and actions to be taken. Customers were instructed to review the Medical Device Removal notice, assure that all users are aware of the notice, confirm receipt of this communication by faxing the attached Customer Response Form to the fax number provided or email to [email protected] and call Terumo CVS Customer Service for a Returned Goods Authorization (RGA) number. For questions call Terumo CVS Customer Service at 1-800-521-2818.

Distribution

USA Nationwide Distribution including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and the countries of: CANADA, CHILE, India, Japan, Singapore, AUSTRALIA, Malaysia, BELGIUM, BELGIUM, BRAZIL, CHILE, BRAZIL, BRAZIL, and Philippines.

Quantity

188,282 Cannulae and 248 Custom Tubing Packs